Product Wiki · Recalls · Health Canada — Medical Devices

Valleylab FT10 FT Series Energy Platform — recall 74682

Health Canada — Medical Devices · Canada · 2023-12-05 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
74682
Date of alert
2023-12-05
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
General and plastic surgery
Product type
Medical devices
Reported risk: Upon insertion of a new (unused) Ligasure™ device, the Valleylab™ FT10 energy platform running software versions 4.0.1, 4.0.2 and 4.0.3 may erroneously indicate that the Ligasure™ device was used previously. When this occurs, the energy platform will display error "e420 usage limit" or error "e416 unknown instrument," and the Ligasure™ device would not be allowed to be used. Recall Start Date: November 20, 2023
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/valleylab-ft10-ft-series-energ

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.