Health Canada — Medical Devices · Canada · 2023-12-01 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: It has been determined that the terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed. Because of this, the cable became disconnected, and arcing occurred. Melting of the cable due to this arcing caused the cable securing parts below it to ignite, resulting in fire at the terminal block. Recall Start Date: November 14, 2023
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.