LINQ II Insertable Cardiac Monitor (ICM) — recall 74622
Health Canada — Medical Devices · Canada · 2023-11-21 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Medtronic is notifying customers of a population of LINQ II Insertable Cardiac Monitors (ICM) that underwent a manufacturing process that may allow for moisture to impact electrode performance. This may create the potential for amplified noise and/or overall signal reduction of the ICM, which may interfere with intended recordings of heart rhythms. Recall start date: November 2, 2023
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.