Infusomat® Space Pump IV Set with Anti-Siphon Valve — recall 74606
Health Canada — Medical Devices · Canada · 2023-11-17 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: In the identified lot number range of administration sets, the tubing outer diameter may vary in size. Additionally, the length of the tubing which contacts the air sensor may be too short resulting in the need to stretch the tubing to properly fit within the Infusomat® space pump. If the tubing is stretched or the tubing outer diameter is too small, this may result in suboptimal coupling between the administration set and the air-in-line sensor dampening the signal of the air-in-line sensor. If the signal is dampened, the Infusomat® space pump may generate false air-in-line alarms. Recall start date: October 23, 2023
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.