NIV Angled Mouthpiece with Notch 22F/15F — recall 74572
Health Canada — Medical Devices · Canada · 2023-11-09 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: The manufacturer of the NIV angled mouthpiece with Notch 22F/15F, advised that devices in lots 32304444, 32305269 and 32306017 were produced with unintended open slits on the side of the mouthpiece body. This allows some of the respiratory gas to escape and thus reduces effectiveness of the related Vivo 45 LS US device. Breas Medical traced these lots and determined that shipments from April 4th, 2023 until July 31st, 2023 could be affected by the issue. Recall Start Date: October 13, 2023
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.