Health Canada — Medical Devices · Canada · 2023-10-16 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Sentinel Ch Spa, the manufacturer of iron for use with Architect and Alinity C Iron Reagent Kit, has identified an issue with the reagents. Some customers have experienced elevated quality control (QC) results as well as elevated patient results. A high recovery of iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of quality control and patient samples results, up to 15% for iron products (Alinity C Iron Reagent ref. 08p3920) to be used on Abbott Alinity instrument and 30% for iron products (iron ref. 6k95-41 and iron ref. 6k95-30) to be used on Abbott Architect instrument. Recall start date: September 29, 2023
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.