Health Canada — Medical Devices · Canada · 2023-10-13 · Type I
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: During a complaint investigation it was reported that the rotation of the Cardiohelp Emergency Drive hand crank was blocked or impaired. In course of internal investigations, it was found that the safety washer causes this issue. The safety washer was originally introduced to the Cardiohelp Emergency Drive in order to lengthen the shaft of the emergency drive by the thickness of the safety washer reducing the axial pressure on the circlip, safety washer and bearing. (serial # specific) Recall start date: October 5, 2023
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.