Product Wiki · Recalls · Health Canada — Medical Devices

Provido, Proveo 8 and M530 OHX Microscopes — recall 74338

Health Canada — Medical Devices · Canada · 2023-09-28 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
74338
Date of alert
2023-09-28
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices - Ear, nose and throat - General and plastic surgery
Product type
Medical devices
Reported risk: Complaint trending revealed an adverse trend of complaints stating that a LMS surgical operating microscope unexpectedly shut down caused by a defective 24VDC power supply which is used inside the surgical operating microscopes Provido, Proveo 8 and M530 OHX. Based on the results of the investigations there is a reasonable probability that the MOSFET in the power supply unit of these microscopes will overheat and become defective. Recall Start Date: August 30, 2023
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/provido-proveo-8-and-m530-ohx-

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.