Product Wiki · Recalls · Health Canada — Medical Devices

Meera Operating Tables — recall 74210

Health Canada — Medical Devices · Canada · 2023-08-25 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
74210
Date of alert
2023-08-25
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Physical medicine
Product type
Medical devices
Reported risk: Under certain conditions, it has been identified that an issue might prevent the device from performing as intended. a specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive) of max. 7 sec. of the Meera OR-Table. Recall start date: August 16, 2023
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/meera-operating-tables

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.