Product Wiki · Recalls · Health Canada — Medical Devices

Olympus Pulmonary Endoscopes — recall 74122

Health Canada — Medical Devices · Canada · 2023-08-15 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
74122
Date of alert
2023-08-15
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Anaesthesiology - Ear, nose and throat - Gastroenterology and urology - General and plastic surgery
Product type
Medical devices
Reported risk: Olympus has received three complaints, including one that was associated with an injury, that upon trying to use an Olympus pulmonary endoscope model with an endotracheal tube, the tip became lodged (entrapped) inside the endotracheal tube connector. As a result of the complaint investigation, it was determined that the scope (bending section) was too large for the endotracheal tube connector. Selection of an endotracheal tube where the size is incompatible with the Olympus airway mobilescope can result in delayed procedures, foreign body obstruction, failure to ventilate the patient, hemorrhage, upper airway tract injury, esophageal perforation, and/or airway mobilescope damage. In an effort to maximize patient safety and mitigate any potential risk to patient health, Olympus is providing the following information related to using an endotracheal tube: • before inserting the airway mobilescope with an endotracheal tube into the patient, slide the tube along the entire length of the airway mobilescope insertion section to confirm that the insertion section of the airway mobilescope can be inserted smoothly into the endotracheal tube. If it cannot be inserted smoothly, the covering material of the bending section of the airway mobilescope or the external surface of the insertion section may be damaged. When using a lubricant, make the above confirmation before applying the lubricant. In the event the endotracheal tube cannot slide smoothly over the insertion section of the airway mobilescope, select another size tube, or inspect scope for damage. Olympus is undertaking this action to provide an IFU addendum which specify the correct maximum scope diameter information. Recall start date: July 28, 2023
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/olympus-pulmonary-endoscopes

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.