Reported risk: Philips has become aware of several situations that can result in a loss of availability of the wireless foot switch during a procedure. Identified issues that may result in the wireless foot switch not being available are: • Loss of bluetooth connection due to interferences from other radio equipment • Battery not fully charged • Battery not holding its charge • Damage in the charger, cable and/or connector of the wireless foot switch Recall start date: July 4, 2023
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/philips-allura-xper-and-azurio

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.