Health Canada — Medical Devices · Canada · 2023-06-20 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: The tip-up fenestrated grasper has two versions on the approved product list (APL): 470347-11 and 470347-12. PN 470347-11 uses machined grips, while PN 470347-12 uses grips that have been manufactured via metal-injection molding (MIM). There have been complaints reported for a broken grip issue against the MIM grips. Grip breakage is observed on both upper and/or lower grips. Recall start date: June 8, 2023
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.