Product Wiki · Recalls · Health Canada — Medical Devices

Eclipse Mobility Scooters — recall 73849

Health Canada — Medical Devices · Canada · 2023-06-19 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
73849
Date of alert
2023-06-19
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Physical medicine
Product type
Medical devices
Reported risk: There is a possibility that a small portion of one seat pin batch under the seat base of the affected mobility scooter(s) was not forged in the right conditions. This may result in insufficient strength and may cause a breakage. In the event of an accident where the seat pin breaks, the occupant may fall and cause an injury. Recall start date: May 18, 2021
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/eclipse-mobility-scooters

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.