Product Wiki · Recalls · Health Canada — Medical Devices

Allen Advance Chest Support With Pad — recall 73738

Health Canada — Medical Devices · Canada · 2023-06-01 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
73738
Date of alert
2023-06-01
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Anaesthesiology
Product type
Medical devices
Reported risk: Baxter Corporation is issuing an urgent medical device recall for the Allen Advance Chest Support to inform customers of the potential for the device to crack where the chest base and prone supports attach to the carbon fiber operating room (or) table rail. Recall start date: May 18, 2023
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/allen-advance-chest-support-pa

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.