Product Wiki · Recalls · Health Canada — Medical Devices

Pushtracker E2 & E3 Utilized With Smartdrive MX2+ Power Assist Device — recall 73692

Health Canada — Medical Devices · Canada · 2023-05-19 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
73692
Date of alert
2023-05-19
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Physical medicine
Product type
Medical devices
Reported risk: Through design and development work, Max Mobility has identified a software issue with the application. When multiple processes are running on the watch's central processing unit, the application may crash unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures. Recall start date: May 2, 2023
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/pushtracker-e2-e3-utilized-sma

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.