Health Canada — Medical Devices · Canada · 2023-05-11 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Philips has identified a hardware issue with a metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the incisive CT system. This component may become detached and make contact with other components located within the incisive CT during rotation. Other components could be damaged due to the contact with the detached component. If this issue occurs, a loud noise will be emitted and the system will shut down. The cover may become displaced if the following sequence occurs: • Damaged component(s) come loose • The damaged component(s) contact other components in the system and cause them to break • A broken component makes contact with the top right gantry cover The displaced cover could create a small gap allowing a fragment of a damaged component to be expelled at a low velocity. The maximum weight of a fragment that could be expelled is 60g (2.12 ounces). Recall start date: May 1, 2023
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.