Agency
Health Canada — Medical Devices
Recall identifier
73501
Date of alert
2023-04-13
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Obstetrics and gynaecology
Product type
Medical devices
Reported risk: Huntleigh Healthcare Ltd. has become aware of a software vulnerability, which manifests itself with the fetal monitor's touch screen freezing. This freeze disables the functions that are being displayed. While the occurrence of this vulnerability is extremely rare and audio remains present throughout the screen freeze, user intervention is required to restart the fetal monitor. Recall start date: April 4, 2023
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/team3-intrapartum-and-team3-an

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.