Product Wiki · Recalls · Health Canada — Medical Devices

Access System - Free T3 Assays — recall 73409

Health Canada — Medical Devices · Canada · 2023-03-29 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
73409
Date of alert
2023-03-29
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Chemistry
Product type
Medical devices
Reported risk: Beckman Coulter has determined that Access Free T3 reagent lot 233968 included an insufficient concentration of blocking reagent. Patient samples tested with Access Free T3 reagent lot 233968 may demonstrate repeatable falsely elevated results, which are caused by insufficiently blocked interferents. Recall start date: March 20, 2023
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/access-system-free-t3-assays

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.