Product Wiki · Recalls · Health Canada — Medical Devices

Heater-Cooler Unit HCU 40 — recall 73341

Health Canada — Medical Devices · Canada · 2023-03-17 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
73341
Date of alert
2023-03-17
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
General hospital and personal use
Product type
Medical devices
Reported risk: Maquet Cardiopulmonary GmbH determined two trends for complaints regarding the Heater Cooler Unit HCU 40: degraded internal drain hoses and leaking vacuum valves. This recall will inform users, and give instructions to replace affected parts during next annual preventive maintenance. Recall start date: March 13, 2023
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/heater-cooler-unit-hcu-40-0

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.