Product Wiki · Recalls · Health Canada — Medical Devices

Permanent Life Support and HLS Sets — recall 72591

Health Canada — Medical Devices · Canada · 2023-01-17 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
72591
Date of alert
2023-01-17
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Cardiovascular - Gastroenterology and urology
Product type
Medical devices
Reported risk: Potentially compromised sterile barrier. The conformity of the products was questioned by the German health authority due to packaging tests not adequately performed. Single sterilized and transport conditioned according to ASTM d4169-22 was conducted. However, the test samples were not double sterilized to cover an assumed worst condition of sterilization impact. These required tests are currently under way, however these test results will not be available until earliest in April 2023. Recall start date: January 9, 2023
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/permanent-life-support-and-hls

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.