Health Canada — Medical Devices · Canada · 2022-12-22 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: There is a potential that Cadd-legacy pumps may not detect that 50 ml and 100 ml Cadd medication cassette reservoirs with flow stop are attached to the pump when the cassettes are properly attached. This issue does not impact 250 ml flow-stop and non-flow stop Cadd medication cassette reservoirs. Manufacturing variations on certain cadd medication cassette reservoirs with flow stop may interfere with the pump detecting a properly attached cadd cassette. In such situations, the Cadd-legacy pump will issue a no disposable attached (NDA) double-beep audible warning if the pump cannot determine that the Cadd cassette is properly attached. The pump will initiate an NDA alarm if the nda double-beep warning is not resolved within 2 minutes. The user must clear the alarm and resolve the cause of the nda event before using the pump. Recall start date: November 29, 2022
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.