Agency
Health Canada — Medical Devices
Recall identifier
72234
Date of alert
2022-12-02
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Anaesthesiology
Product type
Medical devices
Reported risk: Baxter Corporation is communicating two separate, but associated, issues related to Revaclear dialyzers. The first issue is linked to reports of internal blood leaks being caused by a twisted gasket in the dialyzer, prompting Baxter to recall the impacted lots. The second issue is linked to reports of polyurethane displacement in the dialyzer, prompting Baxter to issue a safety alert for the impacted lots. Recall start date: November 24, 2022
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/revaclear-400-dialyzer-revacle

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.