Product Wiki · Recalls · Health Canada — Medical Devices

Aquac Uno H Reverse Osmosis System - Main Unit — recall 72215

Health Canada — Medical Devices · Canada · 2022-11-25 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
72215
Date of alert
2022-11-25
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Chemistry
Product type
Medical devices
Reported risk: The reason for this recall is to inform the operators and users of one Aquac Uno H Reverse Osmosis System of a field safety corrective action due to the possibility of missing ferrite cores in the internal network cable of the device. To meet the requirements of IEC 60601-1 and IEC 60601-1-2 standards, two ferrite cores must be attached to the network cable and the cable routing of the internal network cable must be positioned correctly. Recall start date: November 14, 2022
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/aquac-uno-h-reverse-osmosis-sy

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.