Product Wiki · Recalls · Health Canada — Medical Devices

Easystand Bantam — recall 72189

Health Canada — Medical Devices · Canada · 2022-11-14 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
72189
Date of alert
2022-11-14
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Physical medicine
Product type
Medical devices
Reported risk: Discovered through customer complaint on 9/23/22 and 9/30/22. This issue may not effectively lock the front footrest carrier bracket to the stud on the main frame locking out the supination of that bracket. This results in the unit going into the supine position. Recall start date: October 4, 2022
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/easystand-bantam

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.