Sonicaid Freedom Wireless Fetal Monitoring System — recall 72082
Health Canada — Medical Devices · Canada · 2022-10-26 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Huntleigh has received a report from a user regarding the potential for inappropriate clinical diagnosis and action due to the incorrect setup of the RF channels. In the particular case in question, both units had been inadvertently set to the same RF channel and were placed in close proximity to each other, leading to the second unit displaying the FHR from the first unit. Recall start date: October 19, 2022
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.