Health Canada — Medical Devices · Canada · 2022-08-02 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Ortho Clinical Diagnostics (Ortho) became aware of an anomaly that affects Vitros Xt 3400 and Vitros Xt 7600 systems. These systems utilize a digital imaging reflectometer and led light source to generate Vitros microslide assay results. Ortho's on-going investigation has determined that performing a system shutdown on Vitros Xt 3400 and Vitros Xt 7600 systems may, in some instances, cause a positive or negative shift in the electrical current supplied to the led used in conjunction with the digital imaging reflectometer. This shift in electrical current may also lead to a shift in multiple Vitros microslide assay results reported by the system. Recall start date: July 22, 2022
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.