Hls Set Advanced 5.0 and 7.0 With Bioline Coating — recall 64399
Health Canada — Medical Devices · Canada · 2022-07-25 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Maquet Cardiopulmonary Gmbh (MCP) has received customer complaints of damage to the primary packaging of the Hls Set Advanced. In all it was reported that the Tyvek cover was damaged. Investigation suggest that the Tyvek damage only occurs if the set drops upside down so either the velcro straps, holding the Hls modules, come loose or the Hls module detaches. A shift of the Hls module may cause the venous probe tab(s) to perforate the Tyvek cover. Possible sterility compromise. Recall start date: July 18, 2022.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.