Product Wiki · Recalls · Health Canada — Medical Devices

Alinity and Architect A1C Reagent Kit — recall 64283

Health Canada — Medical Devices · Canada · 2022-06-20 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
64283
Date of alert
2022-06-20
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Chemistry
Product type
Medical devices
Reported risk: This recall is to inform of a potential for falsely elevated hemoglobin %A1C and hemoglobin A1C patient results to be generated using the architect hemoglobin A1C assay when analyzing whole blood or hemolysate samples with poor sample integrity, due to micro-clots and particulate matter, associated with low total hemoglobin values. to address this issue, a revision to the total hemoglobin low linearity value for the architect hemoglobin a1c whole blood and hemolysate applications has been implemented. Recall start date: June 15, 2022
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/alinity-and-architect-a1c-reag

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.