Product Wiki · Recalls · Health Canada — Medical Devices

Evolve Triad Plate Cutter — recall 64258

Health Canada — Medical Devices · Canada · 2022-06-13 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
64258
Date of alert
2022-06-13
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Orthopaedics
Product type
Medical devices
Reported risk: Stryker has identified a nonconformance relating to the Evolve Triad Plate Cutter provided in the Evolve Triad Instrument Kit (4951kit1). Specifically, the plate cutter is unable to meet sterilization requirements when it is sterilized in a double wrapped configuration according to the option 1 steam sterilization parameters detailed in Instructions for Use (IFU150824). While the surgical technique does not require use of the plate cutter, it is provided in the instrument kit so the coronoid plate can be trimmed when needed. Recall start date: June 3, 2022
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/evolve-triad-plate-cutter

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.