Product Wiki · Recalls · Health Canada — Medical Devices

Duraclone B27 Reagent Kit — recall 64235

Health Canada — Medical Devices · Canada · 2022-06-07 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
64235
Date of alert
2022-06-07
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Beckman coulter has discovered that affected Duraclone B27 Reagent Kits may exhibit reduced fluorescence intensity potentially leading to incorrect results. Consequently, Duraclone B27 may not conform to sensitivity and accuracy claim of 100% and 95% respectively as outlined in the product instructions for use (IFU). Recall start date: May 23, 2022
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/duraclone-b27-reagent-kit

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.