Eea Autosuture Circular Stapler With Dst Series — recall 64233
Health Canada — Medical Devices · Canada · 2022-06-07 · Type II
As described by Health Canada — Medical Devices: Eea Autosuture Circular Stapler With Dst Series
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Notification to inform customers regarding potential for the staple guide to not be securely attached to the instrument. A staple guide not attached to the instrument could cause the component to disengage and if disengaged, could allow the device to transect tissue without forming staples. This could result in delay of treatment, extended hospital stay, unspecified tissue injury, unintended radiation exposure, unexpected medical intervention, foreign body in patient, failure to anastomose, and hemorrhage. Recall start date: May 16, 2022
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.