Product Wiki · Recalls · Health Canada — Medical Devices

Pico50 Arterial Blood Sampler — recall 64108

Health Canada — Medical Devices · Canada · 2022-04-25 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
64108
Date of alert
2022-04-25
Country / jurisdiction
Canada
Authority classification
Type II
Notice type
recall
Product category
Anaesthesiology
Product type
Medical devices
Reported risk: Radiometer has been performing integrity testing of the Pico50 arterial blood sampler packaging material. Integrity studies included sterility testing to show that the integrity of the sterile barrier system is intact. In the 24 month real time study, 1 out of 40 samples did not pass the sterility test acceptance criteria. This indicates that the sterile barrier system is compromised after 2 years storage and maintaining the product sterile throughout the shelf life cannot be showed. Recall start date: Apr 13, 2022
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/pico50-arterial-blood-sampler

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.