Health Canada — Medical Devices · Canada · 2021-12-08 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: The affected vitros CKMB slides coating does not adequately inhibit CK-mm up to a total CK more than 1000 u/L. Therefore, samples with a total CK value less than 1000 u/L may exhibit falsely elevated vitros CKMB slides results. Recall start date: Nov 30, 2021
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.