Product Wiki · Recalls · Health Canada — Medical Devices

XP003/20/P Mammomat Revelation AMC Box: Grey Images (2020-04-08) — recall 63404

Health Canada — Medical Devices · Canada · 2020-05-08 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
63404
Date of alert
2020-05-08
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Siemens has identified some issues with the AMC box on the multitom rax unit. The system sporadically produces grey images that are being generated via the one-point technique. The issue is caused by an AMC box (automatic exposure control multic converter). A new revision level of the AMC box is being implemented. It contains a change of the power supply and implementation of an additional pull-down resistor. Siemens is releasing this update instruction to provide new amc box to resolve the issue of sporadically generated grey images.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/xp00320p-mammomat-revelation-a

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.