Product Wiki · Recalls · Health Canada — Medical Devices

Central Water Plant (CWP) Reverse Osmosis 208V Units (2020-03-23) — recall 63324

Health Canada — Medical Devices · Canada · 2020-04-17 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
63324
Date of alert
2020-04-17
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Baxter Corporation is issuing an urgent medical device correction for the Central Water Plant (CWP) Reverse Osmosis 208v units in order to correct devices where there is potential for insufficient crimping of the cables within the main pump. The issue was identified as a result of customer complaints where the top of the main pump had visible burn marks due to heat build-up.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/central-water-plant-cwp-revers

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.