Product Wiki · Recalls · Health Canada — Medical Devices

Azurion 7 M20 (2020-03-03) — recall 63286

Health Canada — Medical Devices · Canada · 2020-04-15 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
63286
Date of alert
2020-04-15
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Philips has identified that no torque was specified for the four screws that connect the gearbox and the mounting flange within the c-arm roll motor. If the torque is too low, the movement of the system may lead to stress on the screws, which potentially could result in the four screws becoming loose. If this occurs, the c-arm could rotate freely over the "roll" movement and the system would not be available for use.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/azurion-7-m20-2020-03-03

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.