Trima Accel Platelet and Multiplasma Sets (2019-12-19) — recall 63179
Health Canada — Medical Devices · Canada · 2020-02-28 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Complaints have been received by Terumo BCT for pressure test error alerts that are associated with inflation and pressurization of the sample bag. A new sample bag line clamp was introduced in early 2019 to mitigate the issue. There has been an increase in the frequency of complaints for the issue since the introduction of the new clamp, including one incident in which air was infused into a donor. Root cause investigation identified two potential causes for the failure of the new sidewall clamp.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.