Heartmate 3 LVAS Implant Kit and Modular Cable With Cap (2019-12-03) — recall 63097
Health Canada — Medical Devices · Canada · 2020-01-24 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Reports received of the HeartMate 3 modular cable being incorrectly inserted into the controller by patients during controller exchange. If the modular cable is incorrectly inserted at 180 degrees from the proper orientation, there is potential for erroneous electrical contact to be made, which may cause a blown fuse, confusion for the patient, and the controller to alarm with " controller fault". In this case, although a physical connection appears to have been made, there is no electrical connection to the power pump.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.