Cobas H 232 Instruments (2019-10-22) — recall 63010
Health Canada — Medical Devices · Canada · 2020-01-06 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Potential defect with the cardiac pipettes used with the cobas H 232 system. Please be advised that this urgent medical device correction (Roche ref. No. rr201929) is not a product removal, as the product itself can still be used as long as users follow the instruction provided in the warning stated in the action taken by Roche diagnostics. In the past months Roche have received 9 customer complaints worldwide for cardiac pipettes (out of more than 11 million distributed single syringes). The following issues were observed: 8 cases where the needle came off the pipette and 1 case where the needle became stuck in a sample tube. The manufacturer has carried out an investigation and confirmed these findings. In all related complaints the issue was detected by the user and no injury occurred.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.