Automated Field Steam Sterilizer (2021-08-19) — recall 62879
Health Canada — Medical Devices · Canada · 2021-10-01 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: FDI has determined through internal testing on august 11th 2021, that the rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground. To be fully compliant with the electrical standard ul 61010-1 2nd edition, section 6.5.2.4, All metal components accessible by the user must have an impedance to ground of 0.1 Ohms or less. In the unlikely event of a damaged or faulty wire within the rear heater assembly, the rear heater plate and cover assembly could potentially become energized.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.