Health Canada — Medical Devices · Canada · 2021-10-01 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Stryker has determined through internal investigation (during testing at the production level) that a fuse block located within the system requires additional insulation to ensure the airo truct complies with applicable electronic requirements for medical devices (iec 60601-1).
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.