Product Wiki · Recalls · Health Canada — Medical Devices

TRACKABLE SUCTION-SUCTION SET (2021-04-22) — recall 62404

Health Canada — Medical Devices · Canada · 2021-05-13 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
62404
Date of alert
2021-05-13
Country / jurisdiction
Canada
Authority classification
Type II
Notice type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: After an in-house inspection on a selection of suction tubes, multiple samples were identified with visible burrs/metal filings affixed along the suction tube inner perimeter. Further investigation has determined that the defect is related to a deficiency in the supplier production process.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/trackable-suction-suction-set-

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.