Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console… — recall 62386

Health Canada — Medical Devices · Canada · 2022-02-08 · Type II

As described by Health Canada — Medical Devices: Medtronic Bio-Console 560 Extracorporeal Blood Pumping Console (2021-03-05)

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
62386
Date of alert
2022-02-08
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Cardiovascular
Product type
Medical devices
Reported risk: Revision D of recall: As a result of eight (8) customer complaints since revision A of this recall, a malfunction with the system controller module used in the Bio-Console 560 was identified. The malfunction is a result of the DC-DC converter and D88 diode failure because of a transient voltage spike in system controller module. This transient voltage spike has the potential to cause dielectric breakdown of the DC-DC converter front-end circuitry resulting in catastrophic failure/excessive current. Through 03-Jan-2022, this issue has resulted in eight (8) complaint occurrences (prior to use) where the instrument or user interface (UI) was not working (blank UI screen), error code 41 and/or smoking/smoking smell prior to use when the device was powered on. No adverse patient effects occurred as a result of these occurrences. Medtronic began using the affected PN in February 2021 to manufacture, service and implement part replacement per Technical Service Update (TSU) MOD0204. All Bio-Console 560 instruments manufactured (January & February 2021) with or had a system controller module installed with the affected system controller are affected by this issue. Additionally, system controller modules that are available for installation by the field service technicians for revision A of this recall are affected. Recall start date: 2022-01-18
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/medtronic-bio-console-560-extr

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.