Health Canada — Medical Devices · Canada · 2021-04-22 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Performing a technical investigation related to a complaint reported from the USA market, Esaote received notice by the manufacturer of the AC/DC adapter (manufacturer XP Power, model AHM180PS19) of the portable ultrasound system MyLabAlpha that, due to the design of the PCB, an unlikely mechanical failure may reduce the system protection against electric shock hazard. This condition can be caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.