Reported risk: The packs listed above contain component 700000172 which is a bacterial filter. DEKRA Certification GmbH (MCP's notified body) issued the decision letter regarding the design dossier review of Customized Tubing Sets (CTPs) under MDD recertification. Due to the insufficient evidence regarding the shelf life of this component, the overall shelflife of the finished tubing packs containing this component must be reduced by one year.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.