EPIQ ULTRASOUND SYSTEM (2021-01-13) — recall 62042
Health Canada — Medical Devices · Canada · 2021-01-29 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Philips recently discovered an issue associated with EPIC ultrasound systems where the control panel arm assembly could have missing or loose screws. In these cases, if undue force, pressure or weight is applied, the control panel mechanism can fail and break off. We are aware of three instances where the control panel arm with missing/loose screws broke and collapsed, one of them causing a minor injury to the operator.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.