Health Canada — Medical Devices · Canada · 2020-11-27 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: The service life of LCSU 4 units manufactured in the period 9 August 2018 to 6 June 2020 may be reduced due to component failures within the internal vacuum air pump. These component failures will result in the LCSU 4 continuing to operate, but not able to provide any suction. As of Sept 25, 2020, Laerdal had received complaints, from less than two percent of all potentially affected LCSU 4s, involving this failure condition. Prolonged use of the LCSU 4 at vacuum levels above 300 mmHg, either during device testing or during patient use, may cause an air pump component to fail resulting in loss of all suction. Alternate suction devices or methods will be required to continue suctioning.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.