Product Wiki · Recalls · Health Canada — Medical Devices

BLOOD MONITORING UNIT 40 (BMU 40) (2020-09-22) — recall 61746

Health Canada — Medical Devices · Canada · 2020-10-16 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
61746
Date of alert
2020-10-16
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: The printed circuit board (PCBA Connector) of the BMU40 was designed with a short creepage distance. This PCBA Connector is positioned in the inner and backside of the metal enclosure of the BMU40. This may lead to shock, loss of data, or delay of use. Maquet Cardiopulmonary GmbH has not received any reports of serious injuries or death due to the malfunction of the PCBA Connector.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/blood-monitoring-unit-40-bmu-4

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.