BLOOD MONITORING UNIT 40 (BMU 40) (2020-09-22) — recall 61746
Health Canada — Medical Devices · Canada · 2020-10-16 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: The printed circuit board (PCBA Connector) of the BMU40 was designed with a short creepage distance. This PCBA Connector is positioned in the inner and backside of the metal enclosure of the BMU40. This may lead to shock, loss of data, or delay of use. Maquet Cardiopulmonary GmbH has not received any reports of serious injuries or death due to the malfunction of the PCBA Connector.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.