MICROSTREAM CO2 EXTENSION MODULE (2020-06-05) — recall 61626
Health Canada — Medical Devices · Canada · 2020-09-25 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Affected M3015A extensions are incompatible with the IntelliVue X3 patient monitor (model no. 367030) and IntelliVue MMX Multi-measurement server. If the extension is connected to an X3 or MMX, which is in turn connected to a host monitor, communication between the X3/MMX and host monitor could be disrupted, resulting in a loss of monitoring and one of the following two technical alarms on the host monitor: < X3 unplugged > < MMX unplugged >
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.