Product Wiki · Recalls · Health Canada — Medical Devices

Symbia T & S Systems (2020-08-20) — recall 61604

Health Canada — Medical Devices · Canada · 2020-09-18 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
61604
Date of alert
2020-09-18
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Siemens has identified that certain Symbia S or Symbia T systems may have unintended slow motion of the detectors in specific situations which could pose potential patient risk. This issue may occur when a detector is moving downwards from approximately the zero-degree position radially towards the center of the camera bore, and an emergency stop press, touch pad engagement or power loss (i.e. system shut down or power failure) results.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/symbia-t-s-systems-2020-08-20

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.